ThermoBLOCK ვარიკოზული ვენების რადიოსიხშირული აბლაციის სისტემა

A radiofrequency ablation platform for incompetent superficial veins, offering thermal closure of saphenous or perforating veins to relieve varicose symptoms.

ის Thermal Coagulation RF Ablation Generator არის მოწინავე, feedback-driven system designed to deliver high-frequency radiofrequency (RF) energy რომ ThermoBLOCK™ catheters და ThermoBLOCK™ stylets—enabling მიზნობრივი thermal ablation for chronic venous insufficiency (CVI). Equipped with a რეალურ დროში monitoring mechanism, the generator precisely regulates intravascular heat parameters, ensuring უსაფრთხო და consistent power application while protecting surrounding tissues.

Real-Time Feedback Control
  • Heat Parameter Monitoring: Intravascular temperature readings and impedance data are continuously analyzed.
  • Automatic Power Regulation: The system modulates RF output based on real-time feedback, maintaining stable thermal therapy without over- or under-heating the vein walls.
  • Targeted Vessel Closure: Concentrates thermal energy along the catheter’s active tip(s), leading to collapsing და sealing of incompetent veins.
  • მინიმალური თანმხლები ზიანი: By adapting energy output to real-time conditions, the generator reduces risk of inadvertent injury to neighboring structures.
  • Digital Touchscreen or Control Panel: Operators can easily set and adjust parameters (power, pulse duration, temperature limits) as needed.
  • Visual & Audible Cues: Real-time displays of energy output, therapy time, and any alerts keep clinicians informed.
  • ThermoBLOCK™ Catheter & Stylet Compatibility: Specifically engineered to function seamlessly with ThermoBLOCK™ components, streamlining venous ablation procedures.
  • Multi-Mode Operation: Adjustable settings accommodate various vein diameters and lesion characteristics, benefiting physicians performing different CVI treatments.
  • Heat Parameter Monitoring: Intravascular temperature readings and impedance data are continuously analyzed.
  • Automatic Power Regulation: The system modulates RF output based on real-time feedback, maintaining stable thermal therapy without over- or under-heating the vein walls.
  • Targeted Vessel Closure: Concentrates thermal energy along the catheter’s active tip(s), leading to collapsing და sealing of incompetent veins.
  • მინიმალური თანმხლები ზიანი: By adapting energy output to real-time conditions, the generator reduces risk of inadvertent injury to neighboring structures.
  • Digital Touchscreen or Control Panel: Operators can easily set and adjust parameters (power, pulse duration, temperature limits) as needed.
  • Visual & Audible Cues: Real-time displays of energy output, therapy time, and any alerts keep clinicians informed.
  • ThermoBLOCK™ Catheter & Stylet Compatibility: Specifically engineered to function seamlessly with ThermoBLOCK™ components, streamlining venous ablation procedures.
  • Multi-Mode Operation: Adjustable settings accommodate various vein diameters and lesion characteristics, benefiting physicians performing different CVI treatments.

პროდუქტის სპეციფიკაციები

სპეციფიკაცია დეტალი / ღირებულება
პროდუქტის დასახელება ThermoBLOCK Varicose Vein RF Ablation System
დანიშნულებისამებრ გამოყენება Endovenous radiofrequency ablation (RFA) of incompetent saphenous or perforator veins
RF Catheter Design 6–7F catheter with electrode segment near distal tip, temperature-controlled
Nominal Power ~10–30 W (depends on vein diameter and system setting)
კათეტერის სიგრძე ~60–120 cm (varies by code)
Generator Compatibility ThermoBLOCK RF Generator or universal 460 kHz system
Temperature Monitoring Integrated thermocouple in ablation segment
Treatment Cycle Typically ~20–40 seconds per segment, auto-regulated by temperature feedback
სტერილობა სტერილური, ერთჯერადი გამოყენების
შენახვის ვადა ~2–3 years at 15–25 °C
კლინიკური ჩვენებები Great saphenous vein (GSV) reflux, small saphenous vein, perforator incompetence
უკუჩვენებები Unclear vein anatomy, severe peripheral arterial disease, known intolerance to device materials

შეკვეთისა და ზომის მატრიცა

კათეტერის ზომა (F) კათეტერის სიგრძე (სმ) Ablation Segment (cm) პროდუქტის კოდი შენიშვნები
6F 60 7 TBLK-6F-60-7 Short catheter for small segments
6F 100 7 TBLK-6F-100-7 Standard ablation length, extended shaft
7F 100 10 TBLK-7F-100-10 Longer ablation segment for large GSV trunk
7F 120 10 TBLK-7F-120-10 For taller patients or more extensive reflux
7F 120 15 TBLK-7F-120-15 Extra ablation segment for major trunk veins